ORIGINAL ARTICLE |
|
Year : 2022 | Volume
: 25
| Issue : 12 | Page : 1973-1977 |
|
An effective artificial testing model to reliability experiments for electronic apex locator devices used in endodontic practice
O Ozdemir
Department of Endodontics, Faculty of Dentistry, Karabük University, Karabük, Turkey
Correspondence Address:
Dr. O Ozdemir Department of Endodontics, Faculty of Dentistry, Karabük University, Karabük - 78050 Turkey
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/njcp.njcp_27_22
|
|
Background and Aim: This experimental study aimed to investigate the reliability of using electronic apex locator devices to determine the working length of artificial root canals. Materials and Methods: The experiments were performed using resin endoblocks and mandibular canine teeth (n = 20/group). After the same working length of root canal samples was provided, the teeth and artificial root canals were embedded in an alginate mold. The measurements with Root ZX® and Propex Pixi® apex locators were performed and recorded. The data were analyzed using SPSS software (SPSS V23; IBM Corp., Armonk, New York, USA) and the variance was set at P < 0.05. Results: There was no significant difference between the groups; mean distance from the actual working length using different apex locators (P = 0.633, P = 0.474), and endpoint positioning distributions (P = 0.591). Conclusion: The results indicate that the artificial model could be a laboratory method of determining the accuracy of apex locators and efficient calibration of devices before their clinical use.
|
|
|
|
[FULL TEXT] [PDF]* |
|
 |
|